Home/EURASIP Journal on Wireless Communications and Networking/Article Record
Peer-Reviewed Article June 22, 2009

On PHY and MAC Performance in Body Sensor Networks

Are you an author of this work?Link to your sovereign AfriScholar profile
Published in: EURASIP Journal on Wireless Communications and Networking
Publisher: Springer Nature • Egypt • 1687-1472
DOI: https://doi.org/10.1155/2009/479512

Abstract

This paper presents an empirical investigation on the performance of body implant communication using radio frequency (RF) technology. In body implant communication, the electrical properties of the body influence the signal propagation in several ways. We use a Perspex body model (30 cm diameter, 80 cm height and 0.5 cm thickness) filled with a liquid that mimics the electrical properties of the basic body tissues. This model is used to observe the effects of body tissue on the RF communication. We observe best performance at 3cm depth inside the liquid. We further present a simulation study of several low-power MAC protocols for an on-body sensor network and discuss the derived results. Also, the traditional preamble-based TMDA protocol is extended towards a beacon-based TDMA protocol in order to avoid preamble collision and to ensure low-power communication.

Download Full Text PDFView at PublisherCrossref RecordCrossref XML

Verified Citation Velocity

Recorded scholarly citations in African and international peer-reviewed literature

74
Total Citations

Citations are continuously tracked and harvested across continental OJS endpoints and OpenAlex feeds.