Home/Computational and Mathematical Methods in Medicine/Article Record
Peer-Reviewed Article December 31, 2010

Reducing False Alarms of Intensive Care Online‐Monitoring Systems: An Evaluation of Two Signal Extraction Algorithms

Are you an author of this work?Link to your sovereign AfriScholar profile
Published in: Computational and Mathematical Methods in Medicine
Publisher: Hindawi Publishing Corporation • Egypt • 1748-670X
DOI: https://doi.org/10.1155/2011/143480

Abstract

Online-monitoring systems in intensive care are affected by a high rate of false threshold alarms. These are caused by irrelevant noise and outliers in the measured time series data. The high false alarm rates can be lowered by separating relevant signals from noise and outliers online, in such a way that signal estimations, instead of raw measurements, are compared to the alarm limits. This paper presents a clinical validation study for two recently developed online signal filters. The filters are based on robust repeated median regression in moving windows of varying width. Validation is done offline using a large annotated reference database. The performance criteria are sensitivity and the proportion of false alarms suppressed by the signal filters.

Download Full Text PDFView at PublisherCrossref RecordCrossref XML

Verified Citation Velocity

Recorded scholarly citations in African and international peer-reviewed literature

63
Total Citations

Citations are continuously tracked and harvested across continental OJS endpoints and OpenAlex feeds.